Ceramic Bearings with Titanium Adapter Sleeves Implanted During Revision Hip Arthroplasty Show Minimal Fretting or Corrosion: a Retrieval Analysis

Volume: 13, Issue: 3, Pages: 241 - 247
Published: Jun 22, 2017
Abstract
BackgroundThe BIOLOX® option system, consisting of a BIOLOX® delta ceramic femoral head with a titanium alloy adapter sleeve, is being increasingly utilized in revision hip arthroplasty. The sleeve protects the ceramic head from fracture and improper motion about the stem trunnion when a damaged trunnion is encountered at revision surgery. Corrosion and fretting due to metal-metal contact at the taper region of hip prosthesis create the...
Paper Details
Title
Ceramic Bearings with Titanium Adapter Sleeves Implanted During Revision Hip Arthroplasty Show Minimal Fretting or Corrosion: a Retrieval Analysis
Published Date
Jun 22, 2017
Volume
13
Issue
3
Pages
241 - 247
Citation AnalysisPro
  • Scinapse’s Top 10 Citation Journals & Affiliations graph reveals the quality and authenticity of citations received by a paper.
  • Discover whether citations have been inflated due to self-citations, or if citations include institutional bias.