PVC membrane ion selective electrode for the determination of pentachlorophenol in water, wood and soil using tetrazolium pentachlorophenolate

Volume: 55, Issue: 3, Pages: 647 - 656
Published: Sep 13, 2001
Abstract
A novel PVC membrane electrode selective for the determination of pentachlorophenolate (PCP) ion based on 2,3,5-triphenyl-2H-tetrazolium (TPT) pentachlorophenolate ion pair as an electroactive material is described. The electrode shows a linear response for PCP ion over the concentration range of 6×10−5 to 1×10−2 M with lower detection limit of 4×10−5 M at 25 °C. The electrode posses a Nernstian slope of −55.0±1.0 mV/decade and a fast potential...
Paper Details
Title
PVC membrane ion selective electrode for the determination of pentachlorophenol in water, wood and soil using tetrazolium pentachlorophenolate
Published Date
Sep 13, 2001
Journal
Volume
55
Issue
3
Pages
647 - 656