Sulfate and sulfide sulfur isotopes (δ 34 S and δ 33 S) measured by solution and laser ablation MC-ICP-MS: An enhanced approach using external correction

Volume: 412, Pages: 99 - 106
Published: Sep 1, 2015
Abstract
Isotope ratio measurements using a multi-collector inductively coupled plasma mass spectrometer (MC-ICP-MS) commonly use standard-sample bracketing with a single isotope standard for mass bias correction for elements with narrow-range isotope systems measured by MC-ICP-MS, e.g. Cu, Fe, Zn, and Hg. However, sulfur (S) isotopic composition (δ34S) in nature can range from at least − 40 to + 40‰, potentially exceeding the ability of standard-sample...