Abstract
Objective: To evaluate the effect of a closed-loop temperature based feedback algorithm on ablative outcomes for pulsed electric field treatments. Methods: A 3D tumor model of glioblastoma was used to assess the impact of 2 μs duration bipolar waveforms on viability following exposure to open and closed-loop protocols. Closed-loop treatments evaluated transient temperature increases of 5, 10, 15, or 22 °C above baseline. Results: The temperature...