Original paper
Electrochemical and biological characterization of thin-film platinum-iridium alloy electrode coatings: a chronic in vivo study
Volume: 17, Issue: 3, Pages: 036012 - 036012
Published: Jun 22, 2020
Abstract
Objective. To evaluate the electrochemical properties, biological response, and surface characterization of an electrodeposited Platinum-Iridium (Pt-Ir) electrode coating on cochlear implants subjected to chronic stimulation in vivo. Approach. Electrochemical impedance spectroscopy (EIS), charge storage capacity (CSC), charge injection limit (CIL), and voltage transient (VT) impedance were measured bench-top before and after implant and in vivo....
Paper Details
Title
Electrochemical and biological characterization of thin-film platinum-iridium alloy electrode coatings: a chronic in vivo study
Published Date
Jun 22, 2020
Volume
17
Issue
3
Pages
036012 - 036012
Citation AnalysisPro
You’ll need to upgrade your plan to Pro
Looking to understand the true influence of a researcher’s work across journals & affiliations?
- Scinapse’s Top 10 Citation Journals & Affiliations graph reveals the quality and authenticity of citations received by a paper.
- Discover whether citations have been inflated due to self-citations, or if citations include institutional bias.
Notes
History