Electrochemical and biological characterization of thin-film platinum-iridium alloy electrode coatings: a chronic in vivo study

Volume: 17, Issue: 3, Pages: 036012 - 036012
Published: Jun 22, 2020
Abstract
Objective. To evaluate the electrochemical properties, biological response, and surface characterization of an electrodeposited Platinum-Iridium (Pt-Ir) electrode coating on cochlear implants subjected to chronic stimulation in vivo. Approach. Electrochemical impedance spectroscopy (EIS), charge storage capacity (CSC), charge injection limit (CIL), and voltage transient (VT) impedance were measured bench-top before and after implant and in vivo....