T1 mapping performance and measurement repeatability: results from the multi-national T1 mapping standardization phantom program (T1MES)
Abstract
Background The T 1 Mapping and Extracellular volume (ECV) Standardization (T1MES) program explored T 1 mapping quality assurance using a purpose-developed phantom with Food and Drug Administration (FDA) and Conformité Européenne (CE) regulatory clearance. We report T 1 measurement repeatability across centers describing sequence, magnet, and vendor performance. Methods Phantoms batch-manufactured in August 2015 underwent 2 years of structural...
Paper Details
Title
T1 mapping performance and measurement repeatability: results from the multi-national T1 mapping standardization phantom program (T1MES)
Published Date
May 7, 2020
Volume
22
Issue
1
Citation AnalysisPro
You’ll need to upgrade your plan to Pro
Looking to understand the true influence of a researcher’s work across journals & affiliations?
- Scinapse’s Top 10 Citation Journals & Affiliations graph reveals the quality and authenticity of citations received by a paper.
- Discover whether citations have been inflated due to self-citations, or if citations include institutional bias.
Notes
History