Test validation, method comparison and reference range for the measurement of β-hydroxybutyrate in peripheral blood samples

Volume: 30, Issue: 1, Pages: 118 - 127
Published: Feb 15, 2020
Abstract
The measurement of β-hydroxybutyrate (βOHB) concentrations is a corner stone of the diagnosis of diabetic ketoacidosis and other ketonic states. The aim of this study was to perform a validation of a peripheral blood βOHB assay (Randox) on a Roche cobas c502 analyser and to establish a βOHB reference range for the validated assay.Precision, linearity and limit of detection and blank (LoD, LoB) were determined according to Clinical and Laboratory...
Paper Details
Title
Test validation, method comparison and reference range for the measurement of β-hydroxybutyrate in peripheral blood samples
Published Date
Feb 15, 2020
Volume
30
Issue
1
Pages
118 - 127