Reliability of the Dexcom G6 Continuous Glucose Monitor During Hyperbaric Oxygen Exposure

Volume: 22, Issue: 5, Pages: 360 - 366
Published: May 1, 2020
Abstract
Background: People with diabetes-related ulcers may benefit from hyperbaric oxygen (HBO2) therapy and from continuous glucose monitors (CGM). Although blood glucose (BG) meters based on glucose oxidase (GO) report erroneously low values at high pO2, BG meters based on glucose dehydrogenase (GD) do not. We therefore examined the performance of a GO-based CGM system in comparison to GO-based and GD-based BG systems in normobaric air (NBAir),...
Paper Details
Title
Reliability of the Dexcom G6 Continuous Glucose Monitor During Hyperbaric Oxygen Exposure
Published Date
May 1, 2020
Volume
22
Issue
5
Pages
360 - 366