How to minimize air embolisms during thoracic endovascular aortic repair with Relay Pro?

Volume: 30, Issue: 2, Pages: 293 - 295
Published: Nov 13, 2019
Abstract
The aim of this study was to evaluate the volume of air released from Relay Pro NBS thoracic stent grafts using different de-airing methods. The gas volume was measured in an in vitro experimental setting. Stent grafts were de-aired via (i) standard saline flushing (40 ml), (ii) increased volume saline flushing (120 ml), (iii) carbon dioxide followed by 40 ml saline flushing and (iv) de-airing with 40 ml of saline in an ultrasound bath. The...
Paper Details
Title
How to minimize air embolisms during thoracic endovascular aortic repair with Relay Pro?
Published Date
Nov 13, 2019
Volume
30
Issue
2
Pages
293 - 295