Apical aortic blood pump preclinical assessment for long‐term use: Durability test and stator topology to reduce wear in the bearing system

Volume: 44, Issue: 8, Pages: 779 - 784
Published: Nov 6, 2019
Abstract
This study presents an assessment for long‐term use of the apical aortic blood pump (AABP), focusing on wear reduction in the bearing system. AABP is a centrifugal left ventricle assist device initially developed for bridge to transplant application. To analyze AABP performance in long‐term applications, a durability test was performed. This test indicated that wear in the lower bearing pivot causes device failure in long‐term. A wear test in...