Development of a small X‐ray source using an ultraviolet laser and pyroelectric crystal

Volume: 48, Issue: 6, Pages: 691 - 695
Published: Jul 30, 2019
Abstract
We investigate a miniaturized X‐ray source using an ultraviolet (UV) laser and a pyroelectric crystal and discuss potential applications in medicine. The UV laser is the fourth harmonic of a Nd:YAG laser with a wavelength of 266 nm, repetition rate of 10 Hz, and pulse energy of 40 mJ. The pyroelectric crystal is a LiNbO 3 cylinder of diameter 10 mm and length 6 mm. The prototype X‐ray source we fabricated is an aluminum parallelepiped of...
Paper Details
Title
Development of a small X‐ray source using an ultraviolet laser and pyroelectric crystal
Published Date
Jul 30, 2019
Volume
48
Issue
6
Pages
691 - 695