Comparative Dose Accuracy of Durable and Patch Insulin Pumps Under Laboratory Conditions

Volume: 21, Issue: 7, Pages: 371 - 378
Published: Jul 1, 2019
Abstract
Background: Recent studies demonstrate variable results of the accuracy with which patch pumps infuse insulin. Aim of this evaluation was to measure dose accuracies of the patch pump mylife™ OmniPod® (OP) in comparison with the durable insulin pump MiniMed® 640G (MM) simulating real-life clinical situations under laboratory conditions. Methods: Thirty-two OP and 15 MM were tested using insulin aspart at five different boluses (0.5, 1, 5, 10, and...
Paper Details
Title
Comparative Dose Accuracy of Durable and Patch Insulin Pumps Under Laboratory Conditions
Published Date
Jul 1, 2019
Volume
21
Issue
7
Pages
371 - 378