Hip simulator testing of the taper‐trunnion junction and bearing surfaces of contemporary metal‐on‐cross‐linked‐polyethylene hip prostheses

Volume: 108, Issue: 1, Pages: 156 - 166
Published: Mar 29, 2019
Abstract
Adverse reaction to metal debris released from the taper‐trunnion junction of modular metal‐on‐polyethylene (MoP) total hip replacements (THRs) is an issue of contemporary concern. Therefore, a hip simulator was used to investigate material loss, if any, at both the articulating and taper‐trunnion surfaces of five 32‐mm metal‐on‐cross‐linked‐polyethylene THRs for 5 million cycles (Mc) with a sixth joint serving as a dynamically loaded soak...
Paper Details
Title
Hip simulator testing of the taper‐trunnion junction and bearing surfaces of contemporary metal‐on‐cross‐linked‐polyethylene hip prostheses
Published Date
Mar 29, 2019
Volume
108
Issue
1
Pages
156 - 166
Citation AnalysisPro
  • Scinapse’s Top 10 Citation Journals & Affiliations graph reveals the quality and authenticity of citations received by a paper.
  • Discover whether citations have been inflated due to self-citations, or if citations include institutional bias.