Phase-Imaging Ion-Cyclotron-Resonance technique at the JYFLTRAP double Penning trap mass spectrometer

Volume: 54, Issue: 9
Published: Sep 1, 2018
Abstract
The Phase-Imaging Ion-Cyclotron-Resonance (PI-ICR) technique has been commissioned at the JYFLTRAP double Penning trap mass spectrometer. This technique is based on projecting the ion motion in the Penning trap onto a position-sensitive multichannel-plate ion detector. Mass measurements of stable 85 Rb ^{+}and 87 Rb ^{+}ions with well-known mass values show that relative uncertainties \Delta m/m \leq 7\cdot 10^{-10}are possible to...