[The Dose Effect of Isocenter Selection during IMRT Dose Verification with the 2D Chamber Array].

Volume: 39, Issue: 3, Pages: 222 - 4
Published: Mar 1, 2015
Abstract
To investigate the dose effect of isocenter difference during IMRT dose verification with the 2D chamber array. The samples collected from 10 patients were respectively designed for IMRT plans, the isocenter of which was independently defined as P(o), P(x) and P(y). P(o) was fixed on the target center and the other points shifted 8cm from the target center in the orientation of x/y. The PTW729 was used for 2D dose verification in the 3 groups...