Reliability and accuracy of six hand-held blood lactate analysers.

Volume: 14, Issue: 1, Pages: 203 - 14
Published: Mar 1, 2015
Abstract
The reliability and accuracy of five portable blood lactate (BLa) analysers (Lactate Pro, Lactate Pro2, Lactate Scout+, Xpress™, and Edge) and one handheld point-of-care analyser (i-STAT) were compared to a criterion (Radiometer ABL90). Two devices of each brand of analyser were assessed using 22 x 6 mL blood samples taken from five subjects at rest and during exercise who generated lactate ranging ~1-23 mM. Each sample was measured...
Paper Details
Title
Reliability and accuracy of six hand-held blood lactate analysers.
Published Date
Mar 1, 2015
Journal
Volume
14
Issue
1
Pages
203 - 14
Citation AnalysisPro
  • Scinapse’s Top 10 Citation Journals & Affiliations graph reveals the quality and authenticity of citations received by a paper.
  • Discover whether citations have been inflated due to self-citations, or if citations include institutional bias.