Frequency-Specific Threshold Determination with the CERAgram Method: Basic Principle and Retrospective Evaluation of Data

Volume: 4, Issue: 1, Pages: 12 - 27
Published: Dec 18, 1998
Abstract
A method for the objective evaluation of the hearing threshold using cortical evoked response audiometry was developed. The method results in a kind of objective audiogram, visualizing the significance of an auditory evoked potential (AEP) in a scheme similar to a conventional audiogram. In the present implementation of the method, four frequencies are tested quasi-simultaneously (500, 1000, 2000 and 3000 Hz; intensity steps of 5 dB). The...
Paper Details
Title
Frequency-Specific Threshold Determination with the CERAgram Method: Basic Principle and Retrospective Evaluation of Data
Published Date
Dec 18, 1998
Volume
4
Issue
1
Pages
12 - 27
Citation AnalysisPro
  • Scinapse’s Top 10 Citation Journals & Affiliations graph reveals the quality and authenticity of citations received by a paper.
  • Discover whether citations have been inflated due to self-citations, or if citations include institutional bias.