Original paper
Electromagnetic interference in a cardiac pacemaker during cauterization with the coagulating, not cutting mode
Volume: 27, Issue: 4, Pages: 527 - 527
Published: Jan 1, 2011
Abstract
Electromagnetic interference in pacemakers has almost always been reported in association with the cutting mode of monopolar electrocautery and rarely in association with the coagulation mode. We report a case of electrocautery-induced electromagnetic interference with a DDDR pacemaker (dual-chamber paced, dual-chamber sensed, dual response to sensing, and rate modulated) in the coagulating and not cutting mode during a spine procedure. We also...
Paper Details
Title
Electromagnetic interference in a cardiac pacemaker during cauterization with the coagulating, not cutting mode
Published Date
Jan 1, 2011
Volume
27
Issue
4
Pages
527 - 527
Citation AnalysisPro
You’ll need to upgrade your plan to Pro
Looking to understand the true influence of a researcher’s work across journals & affiliations?
- Scinapse’s Top 10 Citation Journals & Affiliations graph reveals the quality and authenticity of citations received by a paper.
- Discover whether citations have been inflated due to self-citations, or if citations include institutional bias.
Notes
History