Electromagnetic interference in a cardiac pacemaker during cauterization with the coagulating, not cutting mode

Volume: 27, Issue: 4, Pages: 527 - 527
Published: Jan 1, 2011
Abstract
Electromagnetic interference in pacemakers has almost always been reported in association with the cutting mode of monopolar electrocautery and rarely in association with the coagulation mode. We report a case of electrocautery-induced electromagnetic interference with a DDDR pacemaker (dual-chamber paced, dual-chamber sensed, dual response to sensing, and rate modulated) in the coagulating and not cutting mode during a spine procedure. We also...