Primary migration of a mini-implant under a functional orthodontic loading

Volume: 18, Issue: 3, Pages: 721 - 728
Published: Jul 17, 2013
Abstract
The objective of this study was to examine whether cortical bone thickness and bone mineral density (BMD) can explain the primary migration of a mini-implant under a functional orthodontic tangential loading at the early stage following implantation. Mini-implants were installed in human mandibular sections. A constant tangential load (2 N) was applied to the mini-implant under hydration. Creep, which is a time-dependent viscoelastic...
Paper Details
Title
Primary migration of a mini-implant under a functional orthodontic loading
Published Date
Jul 17, 2013
Volume
18
Issue
3
Pages
721 - 728
Citation AnalysisPro
  • Scinapse’s Top 10 Citation Journals & Affiliations graph reveals the quality and authenticity of citations received by a paper.
  • Discover whether citations have been inflated due to self-citations, or if citations include institutional bias.