A single center comparison between three different apheresis systems for autologous and allogeneic stem cell collections
Abstract
Three different apheresis systems were used in our center for the collection of peripheral blood progenitor cells (PBPCs): COM.TEC (Fresenius Healthcare), COBE Spectra, and Spectra Optia (both from Caridian BCT). We compared 131 autologous and 56 allogeneic apheresis procedures to elucidate feasibility and effectiveness of the different systems. Collection efficiacy varied significantly with lowest results obtained with COBE Spectra. COM.TEC and...