Performance of a fast fiber based UV/Vis multiwavelength detector for the analytical ultracentrifuge
Abstract
The optical setup and the performance of a prototype UV/Vis multiwavelength analytical ultracentrifuge (MWL-AUC) is described and compared to the commercially available Optima XL-A from Beckman Coulter. Slight modifications have been made to the optical path of the MWL-AUC. With respect to wavelength accuracy and radial resolution, the new MWL-AUC is found to be comparable to the existing XL-A. Absorbance accuracy is dependent on the light...
Paper Details
Title
Performance of a fast fiber based UV/Vis multiwavelength detector for the analytical ultracentrifuge
Published Date
Dec 19, 2007
Journal
Volume
286
Issue
2
Pages
121 - 128
Citation AnalysisPro
You’ll need to upgrade your plan to Pro
Looking to understand the true influence of a researcher’s work across journals & affiliations?
- Scinapse’s Top 10 Citation Journals & Affiliations graph reveals the quality and authenticity of citations received by a paper.
- Discover whether citations have been inflated due to self-citations, or if citations include institutional bias.
Notes
History