Determination of 4-hydroxy-2-nonenal in primary rat hepatocyte cultures by liquid chromatography with laser induced fluorescence detection

Volume: 13, Issue: 1, Pages: 75 - 80
Published: Feb 1, 1999
Abstract
An HPLC (high performance liquid chromatography) method with laser induced fluorescence (LIF) detection is described for the determination of 4-hydroxy-2-nonenal (HNE) formed from lipid peroxidation in rat hepatocytes. Carbonyl compounds were fluorescently labelled by incubating the hepatocyte samples with a tagging reagent, 4-(2-carbazoylpyrrolidin-1-yl)-7-nitro-2,1,3-benzoxadiazole (NBD-ProCZ), at 60°C for 10 min. The hydrazone derivatives...
Paper Details
Title
Determination of 4-hydroxy-2-nonenal in primary rat hepatocyte cultures by liquid chromatography with laser induced fluorescence detection
Published Date
Feb 1, 1999
Volume
13
Issue
1
Pages
75 - 80