Ultra-Low-Dose CT of the Thorax Using Iterative Reconstruction: Evaluation of Image Quality and Radiation Dose Reduction

Volume: 204, Issue: 6, Pages: 1197 - 1202
Published: May 22, 2015
Abstract
OBJECTIVE. The purpose of this study is to assess the image quality and radiation dose reduction of ultra-low-dose CT using sinogram-affirmed iterative reconstruction (SAFIRE). SUBJECTS AND METHODS. This prospective study enrolled 25 patients who underwent three consecutive unenhanced CT scans including low-dose CT (120 kVp and 30 mAs) and two ultra-low-dose CT protocols (protocol A, 100 kVp and 20 mAs; protocol B, 80 kVp and 30 mAs) with image...
Paper Details
Title
Ultra-Low-Dose CT of the Thorax Using Iterative Reconstruction: Evaluation of Image Quality and Radiation Dose Reduction
Published Date
May 22, 2015
Volume
204
Issue
6
Pages
1197 - 1202
Citation AnalysisPro